To check a medical device's izin edar (marketing authorisation), open the e-Info Alkes PKRT website at infoalkes.kemkes.go.id, click Cek Izin, search by izin edar number or product name, then match the result against the quotation and read the Status NIE column. Do this before a clinic buys, because medical devices (alat kesehatan, or alkes) and household health supplies (PKRT) may only circulate once they hold an izin edar (Article 50 of Permenkes Nomor 5 Tahun 2026, Minister of Health Regulation 5/2026 on Health Supplies). On the Ministry of Health's official website the check takes a few minutes. Once the device arrives, the clinic also has a recording duty in ASPAK.
This article is for procurement teams, clinic managers, and company HSE staff buying equipment for a site clinic. Calibration and maintenance are covered separately in medical device calibration and maintenance, so we stop here at buying, receiving, and recording. Other company clinic requirements are in permits and management of company clinics. Other regulations in this field are mapped in the occupational health regulation map.
How do you check a medical device's izin edar?
Open the e-Info Alkes PKRT website at infoalkes.kemkes.go.id and click the Cek Izin button on the home page (the Cari menu). The site is run by the Directorate of Medical Device Production and Distribution at the Ministry of Health. You can search by izin edar number, product name, registrant, type, manufacturer, or product kind. Results show the izin edar number, product name, type, registrant, manufacturer, and a Status NIE column. We opened it on 4 November 2026.
Once you find the number, compare the following against the quotation you received:
- The product name and type on the site match the quotation, not merely a similar brand.
- The registrant and manufacturer match the seller or factory named in the documents.
- The Status NIE column shows the izin edar is still valid. We found no explanation of that column's values in the regulations we read, so if the status is anything but clearly valid, ask the Directorate before buying.
- The risk class and product group are plausible for the device. The izin edar standard in Permenkes 11/2025 divides devices into class A (low risk), B (low to moderate), C (moderate to high), and D (high).
- The izin edar number on the packaging or label matches the site. Labelling (penandaan) is governed by Article 52 of Permenkes 5/2026.
The site's menus are split by origin: domestic (Lokal), foreign (Impor), and household health supplies (PKRT) domestic and foreign. In the site addresses these groups use the codes akd, akl, pkd, and pkl. We found no definition of those codes, or of prefixes in izin edar numbers, in Permenkes 11/2025 or 5/2026, so we do not interpret them further. Treat the full number on the site as the only reference.
An izin edar can be revoked. Article 51(1) of Permenkes 5/2026 lists grounds including a harmful product, specifications that differ from the approved data, revocation of the manufacturer's business licence, termination of the distributor or importer appointment, and unmet obligations by the izin edar holder. Article 86(e) and (f) govern the recall of products whose izin edar has ended without renewal or been revoked. So check the same number again if equipment was ordered long ago or in large quantities.
What is the difference between alkes and PKRT?
A medical device is used on people for medical purposes, while PKRT is used for upkeep and care in households and public facilities. The definitions are in Article 1 points 3 and 4 of Permenkes 5/2026, consistent with Article 1 points 12 and 13 of Government Regulation 28 of 2024 (PP 28/2024).
| Alat kesehatan (medical device) | PKRT (household health supplies) | |
|---|---|---|
| Definition | Instruments, apparatus, machines, equipment, implants, in vitro reagents and calibrators, software, and similar materials used on humans for medical purposes that do not achieve their main action by pharmacological, immunological, or metabolic means | Tools, materials, and/or mixtures for upkeep and care that affect human health, intended for households and public facilities |
| Authorisation | Medical device izin edar | PKRT izin edar |
| Validity | The Permenkes 11/2025 standard lists "izin edar validity period" as a data field but does not state its length; read it on the site and in the licence document | At most 5 years, tied to the cooperation letter (PKRT standard in Permenkes 11/2025) |
| Sample procurement question | Is the offered type the same as the registered one? | Is this product really classed as PKRT and not as a medical device? |
The line between the two is not always clear for particular items. Our opinion: if you doubt whether an item is a medical device or PKRT, search for it in both e-Info Alkes menus and keep a screenshot as evidence.
Software also falls within the definition of a medical device. Whether a particular system, such as an electronic medical record application, needs an izin edar is not something we assess here; the record-keeping duties are covered in electronic medical records in clinics.
What are a medical device distributor's duties?
The device distributor formerly called a PAK (Penyalur Alat Kesehatan) is now called a Distributor Alat Kesehatan. Article 125 of Permenkes 5/2026 revokes Permenkes 1191/2010 on the Distribution of Medical Devices, the basis for the old term. The current standard is in the annex to Permenkes 11/2025. Article 63 of Permenkes 5/2026 divides circulation into distribution and handover, and Article 64(1) says pharmaceutical preparations and medical devices may be distributed only by pharmaceutical management facilities, manufacturers, and distributors. Manufacturers and distributors must hold business licences (paragraph (2)), and distribution follows Good Distribution Practice (Cara Distribusi yang Baik, CDB) (paragraph (3)).
For a Distributor Alat Kesehatan, the annex to Permenkes 11/2025 requires, among other things:
| Distributor duty | What it means for the buyer |
|---|---|
| A limited liability company (PT) or cooperative with a business licence (perizinan berusaha, PB) | Ask for the PB number and the legal entity's name |
| A CDB certificate, issued after the PB and before distribution begins | Ask for a copy |
| A full-time technical person in charge (at least D-3) | If the post is vacant, the distributor may not distribute |
| Distributes only devices with an izin edar | Ask for the izin edar number of each product |
| Electronic product listings must show the izin edar number and the PB number | Question any catalogue without them |
| After-sales service for electromedical and in vitro diagnostic devices, plus complaint handling | Ask who handles it and how |
| Routine six-month report through the Health Information System integrated with the National Health Information System | A distributor duty, not a clinic duty |
That standard groups distributor products into five groups: radiation electromedical, non-radiation electromedical, sterile non-electromedical, non-sterile non-electromedical, and in vitro diagnostics. Make sure the groups the distributor covers include the device you are buying. For imported devices that contain or emit ionising radiation or radioactive substances, the import medical device izin edar standard in Permenkes 11/2025 asks for a permit document from the government agency that supervises nuclear energy.
Our opinion: the minimum documents from a supplier are the izin edar number per product, the distributor's PB number and CDB certificate, a manual in Indonesian, a warranty card, and a statement on after-sales service. We suggest keeping these with the inventory.
What are the procurement rules for clinic medical devices?
Article 40(5) of Permenkes 5/2026 says health care facilities obtain health supplies through procurement, grants, or other means. Article 41 prioritises domestic products and domestic content (TKDN). Article 42 requires agencies and health care facilities, government or private, that run national health insurance (JKN) or other health programmes to procure through the electronic catalogue. Paragraph (3) allows other methods where obstacles exist.
Whether your company clinic must use the electronic catalogue is not decided by that article alone; it depends on whether the clinic runs JKN or another health programme. If unsure, ask the dinas kesehatan. The cost of buying equipment is one component of a clinic budget, covered in company clinic cost components.
When goods arrive, Article 49 requires receiving staff to check that the documents are complete against the physical condition, and to record this through the Health Information System integrated with the National Health Information System (paragraphs (2) and (5)). Our opinion: the easiest way is a receiving sheet with the izin edar number, type, serial number, and packaging condition.
Medicines and medical devices have different izin edar and procurement chains. Medicines are not covered here; see medicine management at site clinics. The list of equipment per room that a clinic needs to buy is in medical equipment for a klinik pratama.
What does a clinic record in ASPAK?
ASPAK is a web application that compiles data on facilities, infrastructure, and medical devices at health care facilities, governed by Permenkes Nomor 31 Tahun 2018 (promulgated 1 August 2018). Article 5(1) states that ASPAK must be run by health care facilities, and its annex lists Klinik Pratama and Klinik Utama among the facility types. The main duties:
- Account. A facility must hold an account from the district/city or provincial dinas kesehatan, or the Ministry of Health, according to its licence, applied for using the annex format (Article 6).
- Content. Facility identity, facilities, infrastructure, and medical device data (Article 4(1)). For devices: name, serial number or inventory code, brand, type, year of procurement, year of operation, and calibration status.
- Timing. New data is entered immediately after goods are received, proven by a goods handover record (Berita Acara Serah Terima Barang) (annex). Updates at least twice a year, on 30 June and 31 December, validated by the dinas or the Ministry (Article 7(1) to (3)).
- Officer. The facility leader appoints an ASPAK management officer by decree.
- Removal. A device is removed after a disposal record (Berita Acara Penghapusan) exists.
- Sanctions. Article 10 provides for verbal warnings, written warnings, up to a recommendation to revoke the operating licence. The annex also says the dinas may send a written warning to a facility that does not fill in or update data.
Two limits we should state honestly. First, the ASPAK login page is reachable but cannot be used without an account, so we could not check the application's current screens and fields. Second, the annex to Permenkes 31/2018 refers to Permenkes 54/2015 for devices that must be calibrated, and Permenkes 54/2015 has been revoked (see the calibration article). We found no replacement in the ASPAK text. Permenkes 31/2018 itself is marked In Force on JDIH and is not on the list of regulations revoked by Permenkes 11/2025 or 5/2026, but we did not verify how it is enforced against clinics. Ask your dinas kesehatan.
The table below links purchase documents to ASPAK entries and the clinic inventory. The third column is our practical advice, not an official field.
| Document when the device arrives | Used for | Kept with |
|---|---|---|
| Goods handover record (Berita Acara Serah Terima Barang) | Trigger for entering new data in ASPAK | Clinic inventory |
| Serial number, brand, type from the physical device | ASPAK device data fields | Receiving sheet |
| Copy of the izin edar number and an e-Info Alkes screenshot | Proof of izin edar at purchase | Procurement file |
| Year of procurement and year of operation | ASPAK fields | Inventory |
| Calibration and function test results | Calibration status in ASPAK and maintenance schedule | Maintenance card |
What about calibration and function testing after delivery?
We already covered this in medical device calibration and maintenance: validity of IPAFK certificates and labels, maintenance under Permenkes 15/2023, and scheduling at remote sites. Only two points connect to purchasing. First, Article 3 of Permenkes 15/2023 puts inventory (inventarisasi) as the first activity, so recording in ASPAK and in the clinic inventory serve the same activity. Second, a device bought from a distributor should be function-tested before use, with the result recorded at handover. Well-kept purchase and recording files also help when a clinic prepares for accreditation, which is covered in accrediting a company primary clinic.
An example of equipment whose izin edar and supplier documents should be checked before it enters the clinic is the AED, whose routine checks we cover separately in AEDs in the workplace.
What if the clinic imports equipment itself?
Importing medical devices may only be done by business actors holding a business licence (Article 420(2) of PP 28/2024; Article 57 of Permenkes 5/2026 also names research institutions, members of the public, and other institutions). An imported medical device must have an izin edar (Article 421 of PP 28/2024; Article 58(4) of Permenkes 5/2026), and health supplies are imported through an electronic system integrated with the Indonesia National Single Window (Article 58(1)).
There are limited exceptions. Article 420(4) of PP 28/2024 allows hospitals, clinics, non-profit organisations, and individuals to import pharmaceutical preparations and medical devices for certain circumstances, but paragraph (5) forbids circulating those goods. Article 62 of Permenkes 5/2026 sets a special route: in certain circumstances health supplies can be imported without an izin edar with the Minister's approval, and medical devices and PKRT on this route are used for health services, research, government programmes, charity, exhibitions, personal use, or other needs set by the Minister (paragraphs (1), (3), and (5)). Paragraph (6) exempts imports for personal use and other needs from the Minister's approval. Article 58(5) also lets the Minister issue an import statement (surat keterangan impor) in certain circumstances, and paragraph (6) covers used medical devices for certain purposes after meeting safety, benefit, and quality requirements.
We do not detail the special-route procedure and documents, because the conditions are set by the Minister and we could not find them in full in the texts we read. Our copy of PP 28/2024 is a PDF with low text quality. We read the article numbers carefully, but check them against the text on peraturan.go.id before quoting them in official documents. Our opinion: for a company clinic, the simplest route is to buy from a distributor that holds, or is appointed by, the izin edar holder.
Checklist before buying medical devices
- The product's izin edar number is found in e-Info Alkes and a screenshot is saved.
- Product name, type, registrant, and manufacturer match the quotation.
- The Status NIE and the izin edar validity period are clear.
- The distributor's PB number and CDB certificate have been requested.
- The distributor's product groups cover the device being bought.
- The manual, warranty card, and after-sales statement have been requested.
- Receiving uses a handover record and a sheet checking documents against the physical goods.
- Device data is entered in ASPAK right after receipt, and before the 30 June or 31 December cut-off.
- The device is function-tested, scheduled for calibration, and recorded in the inventory.
Common mistakes
- Checking the brand instead of the izin edar number. One brand can have several types with different authorisations.
- Using the term PAK. Its basis, Permenkes 1191/2010, has been revoked by Permenkes 5/2026.
- Postponing ASPAK entry until the end of the semester. The rule asks for new data to be entered immediately after goods are received.
- Treating an izin edar as proof of fitness for use. An izin edar allows a product to circulate; calibration and function testing are still needed.
- Mixing self-import with ordinary purchasing. Goods imported for certain circumstances may not be circulated (Article 420(5) of PP 28/2024).
Our medical equipment service covers procurement to need, requesting supporting documents from suppliers such as the izin edar, manual, and warranty card and keeping them with the inventory, installation and function testing, and handover recorded with serial number, location, function test result, and recipient.